Flashpoint
 
 MiniFlash FLA/H
 

With the flash point samplers MINIFLASH FLA and FLAH, the manipulation time for 8 different samples is less than 2 minutes.
 

 

Features


‧Fully Automatic 8 Position Sampler
‧4 minutes Manipulation Time for 8 Samples
‧Continuous Operation - up to 80 Samples per Day!
‧Complies with the latest flashpoint standard ASTM D 7094 - MCCCFP (Flash-Point by Modified Continuously Closed Cup) and ASTM D6450 - CCCFP (Flash-Point by Continuously Closed Cup)
‧Excellent Correlation to ASTM D93, D56, ISO2719 (EN/DIN 22719), ISO13736, IP170
‧Maximum Safety through Ignition Protection Technology™: only 1-2 ml sample, no open flame, continuously closed cup, no offensive vapors in the lab
‧Approved by: US D.O.T. and RCRA
‧Approved by: NATO; 6630-25-145-3256, 6630-25-146-0895, 6630-01-534-1774 and the US NAVY - NSN: 6625-01-472-6783
‧Approved for Diesel Fuel and Kerosene specifications ASTM D975, D396, D2880, D3699
‧Approved for the ASTM specifications of recycled fuel and biodiesel
‧The world wide standard in the flavors and fragrances industry
‧The highest precision: details see below details (Precision data)
‧Short measuring time (depends on sample type) due to small sample volume of 1-2 ml
‧Fast thermoelectric temperature regulation by Peltier Protection Technology™
‧Easiest handling (instrument-user dialogue) and fast cleaning of the sample cup
‧Wide Measuring Range: 0 to 400°C (32 to 750°F)
‧Compact & portable for field use (car battery connector)
‧RS232 interface and MINIWIN software (for PC)


Technical Data

‧Recommended Temperature Range
   MINIFLASH FLA    0 to 200°C (32 to 390°F)
   MINIFLASH FLAH  10 to 400°C (50 to 750°F)
   MINIFLASH FLPH   10 to 400°C (50 to 750°F)
‧Power requirements: 100/110/230 V AC 50/60 Hz, 150 W
‧Field application:  12 V/8A DC (vehicle battery)
‧W x H x D: 312 x 402 x 390 mm (12.3" x 15.8" x 15.3")

Precision Data

MINIFLASH
ASTMD6450

MINIFLASH
ASTMD7094

Pensky-Martens
ASTM D93A

Repeatability

1.9°C (3.4°F)

3.8°C (6.8°F)

4.1°C (7.4°F)

Reproducibility

3.1°C (5.6°F)

5.4°C (9.7°F)

6.9°C (13.4°F)

 